Please provide an overview of the Cloe Screen CHI assay.
CHI is an industry recognised high throughput method for determining lipophilicity. This method was originally developed by researchers at GlaxoWellcome13, and is based on reverse phase LC retention times. The method uses a literature accepted series of reference compounds with known CHI values14 to create a calibration line allowing CHI values for test compounds to be calculated according to their retention time.
|
Reference Compound |
CHI |
|
Theophylline |
18.4 |
|
5-Phenyl-1H-tetrazole |
23.6 |
|
Benzimidazole |
34.3 |
|
Colchicine |
43.9 |
|
8 Phenyl-theophylline |
51.7 |
|
Acetophenone |
64.1 |
|
Propiophenone |
77.4 |
|
Butyrophenone |
87.3 |
|
Valerophenone |
96.4 |
Table 2
Table displaying the CHI values
for the reference compounds.
Figure 3
Graph showing the relationship between retention time and literature CHI values.
The gradient is calibrated by determining the retention time of each reference compound on a reverse phase LC system at pH 7.4. The test compounds are analysed on the same system. The equation generated from the calibration line for the reference compounds is then used to determine the unknown CHI value for the test compound.
What is the relationship between LogD and CHI?
CHI correlates closely with LogD7.4 octanol/water partition coefficients as demonstrated in Figure 4.
Figure 4
The data below shows the relationship between LogD7.4 and CHI for a set of 15 unrelated compounds.
How does lipophilicity influence the pharmacokinetic properties of a drug?
Lipophilicity is a key determinant of the pharmacokinetic behaviour of drugs. It can influence distribution into tissues, absorption and the binding characteristics of a drug, as well as being an important factor in determining the solubility of a compound.
How can CHI be used in predicting pharmacokinetics?
Lipophilicity significantly influences ADME properties and is widely used in drug discovery for QSAR (quantitative structure activity relationship) and QSPR (quantitative structure property relationship). This information is often used by our clients to rank compounds and determine which compounds to investigate in further studies.
What range of CHI values should I expect to see with my compound?
As CHI is a measure of retention of the test compound on the LC column a large range of values can be obtained. For compounds that are not retained on the column for any significant length of time a CHI value of <0 will be returned signifying very low lipophilicity. For compounds which are not easily washed off the column a CHI value of >120 is returned signifying very high lipophilicity. For any values in between the actual CHI value will be returned based on the calibration line of the reference compounds.